USES:
Suitable for dental, eye, operating room, supply room and other medical and health department, for all of the packing or without packing solid instrument, class A cavity instruments (dental, endoscopic), implantable devices, accessories fabric, rubber catheter items such as sterilization.
The order no.
|
XG1.UCD-60M
|
XG1.UCD-80M
|
XG1.UCD-135M
|
XG1.UCD-185M
|
XG1.UCD-300M
|
Design pressure
|
-0.1/0.28Mpa
|
The highest working pressure
|
0.23Mpa
|
Vacuum lower limit
|
-0.080Mpa
|
Sterilization temperature
|
105-136℃
|
Temperature display precision
|
0.1℃
|
Pressure display precision
|
1Kpa
|
Cavity material
|
06Cr19Ni10
|
volume
|
60L
|
80L
|
135L
|
185L
|
300L
|
Within the cavity size(φL)
|
380×510mm
|
380×710mm
|
500×710mm
|
500×950mm
|
632×1000mm
|
Overall dimensions(L×W×H)
|
1000×697×1600mm
|
1000×697×1600mm
|
1110×750×1742mm
|
1350×750×1742mm
|
1400×890×1780mm
|
The net weight
|
230Kg
|
260Kg
|
400Kg
|
460Kg
|
750Kg
|
voltage
|
380V
|
380V
|
380V
|
380V
|
380V
|
Power frequency
|
50Hz
|
50Hz
|
50Hz
|
50Hz
|
50Hz
|
power
|
10Kw
|
11Kw
|
11Kw
|
11Kw
|
17Kw
|
The market reference price
|
68200
|
70400
|
90200
|
95700
|
106000
|
Product features:
1, temperature, pressure, time, running state and fault alarm display
2, water injection, pulsating vacuum exhaust, vacuum drying, heating, sterilization, whole process automatically
3, with fabric, instrument, liquid, and rubber, custom and so on sterilization procedures and B&D test, vacuum test
4, the built-in import vacuum pumps, pulsating vacuum for many times, in the dressing, cavity equipment of cold air out completely, to ensure that the steam saturation
5, built-in steam generator
6, sterilizer adopts double cylinder structure, lining for high quality stainless steel
7, after pulsing vacuum drying system, instrument residual humidity no more than 0.2% after drying, dry dressings residual humidity no more than 1%
8, 0.22 mu m aseptic high efficiency air filter, guaranteed to break the vacuum air aseptic air, sterile items to avoid pollution again
9, configure printers, print date, time, procedure parameters
Safety device:
1, automatic overtemperature protection device: pot body temperature more than set temperature, the system automatically cut off the heating power, and displays a report to the police.
2, multiple control and protection: steam generator water level protection, pressure control, overpressure automatic pressure release.
3, safety interlock door: the electronic door lock, only door closed in place, steam sterilizer can be connected into the booster, the program can start;With the pressure, or did not plug in, open the door mechanism be locked, can't open the door.Chamber liquid temperature more than security, unable to open the door, ensure the safety of personnel and items, to avoid splash even bottle.
4, double overpressure protection: automatic pressure relief valve, pressure more than design pressure relief valve open release pressure;Overpressure automatic protection control, pressure more than setting, automatic exhaust steam pressure, according to the police.
5, electronic circuit safety device: the main loop of the communication with short circuit protector, direct current control loop with overpressure overload protection.
|